Stereo microscope — VEVOR JH-1 series

Keep it covered when not in use, and keep the bench under it clear. Support the head with one hand before loosening any clamp.

Stereo microscope on an articulated arm, for inspecting small parts and doing fine work such as engraving, watch repair and small assemblies. Stays at the Precision Corner in the Metal room.

In use

Covered

Stereo microscope at the Precision Corner

Stereo microscope at the Precision Corner


Microscope under its blue cover

Microscope under its blue cover




Location

Precision Corner (Precisionshörnan), Metal room

Note

Stays in the Metal room — do not move it elsewhere

Magnification

7–45× (3.5–22.5× with the 0.5× lens)

Help / faults

Slack #metallrummet

Not a measuring instrument — use calipers or other calibrated tools for dimensions.

Controls

Letters refer to the manual's figures (PDF pp. 6, 9–10).

Control

Function

Knob on the optical head

Zoom

Large round knob on the front arm (D)

Focus

Upper handle (C)

Head angle — support the head while adjusting

Column lock (E)

Arm height

Friction knobs (A, B, F)

Arm stiffness — clockwise = stiffer

T-knobs on forehead rest

Side knob = angle, top knob = depth

Screws (G)

Arm rigidity — maintainers only

How to use

  1. Remove the cover and put it somewhere clean. Clear the bench and place the part in a stable holder.

  2. Position the head over the work, about 165 mm above it.

  3. Adjust eyepiece spacing until you see one image. Adjust the forehead rest.

  4. Focus with knob D, zoom with the head knob. Find the work at low zoom, then zoom in.

  5. Refocus when the work height changes.

Setting the eyepieces for your eyes

  1. Zoom to max, both diopter rings at 0.

  2. Look through the right eyepiece and focus with knob D.

  3. Zoom to min. Without touching D, sharpen with the right diopter ring.

  4. Repeat steps 1–3 until the image stays sharp across the whole zoom range.

  5. At min zoom, sharpen the left eye with the left diopter ring.

When finished

  • Turn off the light, park the arm and put the cover on (it must cover both the objective and the eyepieces).

  • Remove your parts and tools and clean the bench.

  • Don't grind, wire-brush or make chips near the uncovered microscope.

Troubleshooting

Problem

Action

Head drifts or arm won't hold position

Support the head, stop using it, report in #metallrummet.

Blurry, double or shaded image

Check eyepiece spacing and focus. Don't take the head apart.

Dirty lens, damaged cover, loose eyepiece

Report in #metallrummet.

Care

  • Don't touch the lenses. Remove dust with a blower bulb or lens brush only — no rags, paper towels or compressed air.

  • Oily or stubborn dirt: ask an experienced member.

  • Maintainers check the setting screws regularly (manual p. 12).

  • Maintenance and repairs log

Specifications



Model

VEVOR stereo microscope kit, JH-1 series (art. no. 010191167000)

Stand serial no.

241210 5444 G 0002

Eyepieces

WF10X/20 (10×)

Zoom

0.7–4.5×

Total magnification

7–45×; 3.5–22.5× with 0.5× auxiliary lens

Auxiliary lens

0.5×, ~165 mm working distance

Stand

Spring-balanced articulated arm on bench column, forehead rest

Stand label with serial number

Stand label with serial number